K-1430D Ammonia

K-1430D Ammonia HR CHEMets Kit was designed to obtain Ammonia levels from 0-2000 ppm and 0-10,000 ppm

Replacement ampoules for the K-1430D Test Kit can be found here.

Ammonia Information

2-5 Business Days
SKU
K-1430D
  $87.00

Description / K-1430D Ammonia

Why Choose HR CHEMets kits?

HR CHEMets® are the new way to test at high ranges and replaces the now discontinued VACUettes product line. This product is the closest replacement for the discontinued K-1420D test kit.

The K-1430D Ammonia HR CHEMets® Test Kit (HBA) comes in a plastic case and contains everything needed to perform 30 tests (except distilled water): 30 HR CHEMets® ampoules, a color comparator (C-1404), syringe (A-0063), MiniPet® (A-0193), stabilizer solution (A-1404), catalyzer solution (A-1405) , activator solution (A-1410), pipettor tips pack (A-1410), 25 mL sample cup (A-1410), and instructions.

This test kit provides direct read results using a visual color comparison for applications where concentrations of ammonia between 0-2,000 ppm may be present such as wastewater.

What's Included:

     - 30 CHEMets Ampoules

     - Color Comparator 

     - 1 25 mL Sample Cup

     - Stabilizer Solution

     - Catalyzer Solution

     - Activator Solution

     - 3 mL syringe

     - MiniPet

     - Pipettos Tips Pack

     - Instruction Manual

Method:

The K-1430D Ammonia Kit was designed to accurately measure the levels of Ammonia in water from 0-2000 ppm and 0-10,000 ppm. K-1430D uses the Hydroxybenzyl Alcohol (HBA) Method. In the ammonia test method that employs the Hydroxybenzyl Alcohol chemistry, free ammonia reacts with hypochlorite to form monochloramine. Monochloramine reacts with HBA, in the presence of sodium nitroferricyanide, to form a green colored complex.  This test method measures the sum of free ammonia and monochloramine. Results are expressed in ppm (mg/L) ammonia-nitrogen, NH3-N. The Hydroxybenzyl Alcohol Method offers sensitivity similar to the Nesslerization Method and there is no generation of mercury-containing waste.

 

1. Place a pipette tip firmly onto the end of the MiniPet®4 (fig. 1). Note: Use a fresh pipette tip for each test.  2. Depress the plunger on the minipet. Immerse the tip in the sample to be tested and release the plunger. A portion of the sample will be drawn into the tip (fig. 2).
3. Hold the minipet over the empty sample cup, and depress the plunger to dispense sample (fig. 3). 4. Dilute the contents of the sample cup to the 25 ml mark with distilled water (fig. 4). 

5. Add solutions to the sample cup in the following order (fig. 5).

5 drops S-1404 Stabilizer

5 drops S-1405 Catalyzer (green)

5 drops S-1406 Activator (blue)

6. Immediately place the CHEMet ampoule, tip first, into the sample cup. Snap the tip. The ampoule will fill leaving a bubble for mixing (fig. 6).

To mix the ampoule, invert it several times, allowing the bubble to travel from end to end. Dry the ampoule. Obtain a test result  5 minutes after snapping the tip.  

7. Obtain a test result by placing the ampoule between the color standards until the best color match is found (fig. 7).

NOTE: Use the concentration scale on the comparator label that corresponds to the range of the test procedure being used.

0 - 10,000 ppm Test Procedure

  1. Using the syringe provided, obtain 3 mL of the sample to be tested and dispense it into the empty Sample Prep Cup.
  2. 2. Dilute to the 15 mL mark with distilled water. Perform the entire 0 - 2000 ppm Test Procedure using this pre-diluted sample.

For more detailed instructions, see the link at the bottom of the page.

More CHEMetrics Kits for Ammonia:

Vacu-Vials
Product Number Method   Range
K1523 Direct Nesslerization   0-14.0ppm
K1503 Direct Nesslerization   0-7.0ppm
K1413 Hydroxybenzyl Alcohol (HBA)   0-3.00 & 0-60.0ppm
Chemets Kits
Product Number Method MDL Range
K1510 Direct Nesslerization 0.05 ppm 0-1 & 1-10ppm
K1420 Hydroxybenzyl Alcohol (HBA) 0.125 ppm 0-4 & 0-80ppm
Vacuettes
Product Number Method MDL Range
K1520A Direct Nesslerization 25 ppm 25-250 & 125-1250ppm
K1520B Direct Nesslerization 25 ppm 25-250 & 125-1250ppm
K1520C Direct Nesslerization 120 ppm 125-1250 & 1000-10,000ppm
K1520D Direct Nesslerization 5 ppm 5-50 & 25-250ppm
K1420B Hydroxybenzyl Alcohol(HBA) 25 ppm 25-250 & 125-1250ppm
K1430D Hydroxybenzyl Alcohol(HBA) 2000 ppm 0-2000 & 0-10,000ppm

 

Downloads and Links:

K-1430D Instructions

Safety Data Sheet

Technical Information

Ammonia Information

More Information

Manufacturer CHEMetrics
Gas Ammonia (NH3)
4-20 mA Output No
Voltage Output None
Datalogging No
Warranty 1 Year
Method HR CHEMets
MDL 0-2000 ppm
Range 0-2000 ppm, 0-10,000 ppm