UV-106-M Ozone Analyzer
Medium range precision UV Ozone Analyzer. Range 0-1,000 ppm, with resolution of 0.01 ppm resolution allows for medium range ozone measurement up to 1,000 ppm with great accuracy in a compact, easy to use package.
Rental Option HERE
- SKU
- uv-106m
- Size
- 95 × 210 × 216 mm (3.75 × 8.25 × 8.50 in)
Description / UV-106-M Ozone Analyzer
The UV-106-M is a low range ozone analyzer featuring ozone measurements up to 1,000 ppm with low range measurement of ozone down to 0.01 ppm.
Bench mount ozone analyzer is compact and easy to use. The UV-106-M can be purchased in either the bench mount or NEMA enclosure.
Integrated sample pump will pull a sample of ozone from your process or location into the analyzer and measure ozone in that gas stream. A UV lamp and receptor are used to determine the quantity of ozone in gas by determining the rate of ozone absorption by the UV lamp. This method of ozone detection will offer accurate detection of ozone at levels up to 1,000 without compromising low range detection below 1 ppm, while eliminating most cross-sensitivity from other gasses.
If you're looking to temporarily monitor ozone, we have a UV-106-M Ozone Analyzer Rental available here.
Features:
- Measurement based on UV absorption
- Low power consumption; can be battery operated
- Internal data logger with real time clock
- On-board microprocessor with interactive menus includes data averaging options of 10 s, 1 m, 5 m and 1 hr
- USB and RS-232 output of time/date, O3 concentration, internal temperature and pressure
- Analog output (0-2.5 V and 4-20 mA) of ozone concentration in user selected units
- Two level relay for control purposes (e.g., control of ozone source or turn warning light on and off)
- Long-life pump (15,000 hr)
The UV-106-L ozone analyzer can also come with this carrying case. While most ozone analyzers are difficult to transport from place to place, this carrying case makes it extremely easy to move your UV-106-L in between locations. One of it's great features is that the carrying case not only holds your UV-106-L analyzer, but it can also conveniently hold all of the additional cables and tubing. With expensive equipment like this, it is nice to not have to worry about your ozone analyzer incurring any damage when moving from spot to spot.
We highly recommend replacing the filter frequently to prevent contamination of the flow path by particulate matter.
More Information
Manufacturer | 2B Technologies |
---|---|
Quickbooks Stock Level | 2.000000 |
Gas | Ozone (O3) |
Resolution | 0.01 ppm |
Response Time | 10 seconds |
Sensor type | UV Analzyer |
Sampling Method | Fan |
Alarms | Relays |
Power Requirements | 12 VDC with supplied adapter |
Size | 95 × 210 × 216 mm (3.75 × 8.25 × 8.50 in) |
Field Replaceable Battery | No |
Warranty | 1 Year |
Audible Alarm | No |
Calibration | Annually |
Range | 0-1,000 ppm |
Unit Features | - Standard enclosure - Includes datalogging with software - Low power consumption; can be battery operated -15,000 Hr Pump Life -Two Level Relay Control |
Sensor Range | 0-1,000 ppm (O3) |